Digital
indicator with 50 dot bar meter display CSD-531 |
 |
 |
|
-
|
 |
|
| E |
|
50 dot bar meter display with the digital
display. |
| E |
|
Setting value of 2 points of upper limit,
one point of upper/lower limit, 2 points
of lower limit is available. |
| E |
|
Automatic calibration by registering
the rated output data (X.XXX mV/V) of
strain gage applied transducers is performed.
|
| E |
|
Power supply : AC100 V to AC240 V |
|
|
 |
|
| Bridge power supply |
DC5 V}0.25 V within 60 mA
(Changeable to DC2.5 Vj |
| Input range |
F.S. setting can be
made at the input range from }0.4 mV/V to }3.1 mV/V
iAt the time of DC5 V for bridge power supply.j |
| Output range |
DC4 mA to 20 mA
(At
the time of 0.4 mV/V for the bridge power supply)
Load resistance 260 Ά or less |
| Output times |
Changeable to 4 times/s, 20 times/s or 50 times/s
(Synchronized with display
rate) |
| Non-linearity |
0.1 %F.S. |
| Temperature efficient |
Zero point }0.5 ΚV/,
Sensitivity }0.02 %F.S./ (Input conversion) |
| Input noice |
}0.6 ΚVp-p or less
(At the default setting for digital filter and stabilized filter) |
| A/D sampling |
100 times/s |
| Load display |
|
| Display range |
-9 999 to {9 999 (Changeable to x 2,
x 5 or x 10) |
| Bar meter display |
Red or Green 50 dots
bar meter displays the ratio of the current load
value comparing with the maximum display value,
or the ratio of the set value of the comparator
S1 or S2 comparing with the maximum display value. |
| Display times |
4 times/s (Changeable to 20 times/s or 50
times/s) |
| External control function |
ZERO, HOLD |
| Comparator function |
|
| Set value |
-9 999 to {9 999 |
| Set quantity |
2 points of S1and S2 |
| Comparator conversion rate |
Changeable to 4 times/s, 20 times/s or 50 times/s
(Synchronous with display rate) |
| Contact output function |
S1, S2 Output the
contact when less than/or more than the comparator
set value is obtained.
S0 Short of contact at the time of whichever condition
as follows is set.
- When the condition is full(100 % of rated load)
- When the load is between the set value of S1 and
S2 |
| contact output |
1a contact
AC125 V : 3 A (Load resistance)
DC30 V : 3 A (Load resistance) |
| Operating temp. /humidity range |
0 to
50 / 85 %RH or less (Non condensing) |
| Power supply voltage |
AC100 V to 240 V 50/60 Hz
(Permitted changeable range from AC85 V to 264 V)
|
| Outline dimensions(WxHxD) |
110 mm x 110 mm x 205
mm (excludes protruding parts) |
| Weight |
Approx. 1.4 kg (without options) |
|
 |
|
 |
|
| CAD files (DXF) |
Specification sheet (PDF) |
Instruction manual (PDF) |
| CSD-531 (354KB) |
No.EN382531 (73KB) |
No.294-1114 (2,073KB) |
Adobe Acrobat Reader is required to refer the Specification data sheet.
Acrobat Reader is a registered trademark of Adobe Systems Ltd.
The download of Adobe Acrobat Reader is here. |
|
| |